Crude Protein Content

Crude Protein Content digunakan untuk menentukan kandungan nitrogen total dalam berbagai sampel, termasuk sampel padat, semi-padat, dan cair. Alat ini juga dapat digunakan untuk menentukan kandungan protein dalam sampel.

Aplikasi utama Crude Protein Content:

  1. Pertanian: Penentuan nitrogen dalam pupuk, tanah, benih, dan pakan ternak.
  2. Industri makanan: Penentuan kandungan protein dalam daging, produk daging, produk susu, anggur merah, dan sereal.
  3. Industri lingkungan: Penentuan nitrogen dalam limbah cair dan gas industri.
  4. Industri farmasi: Penentuan nitrogen dalam bahan baku farmasi dan obat-obatan.
  5. Penelitian: Penentuan kandungan nitrogen dalam berbagai sampel, seperti sampel air, tanah, dan jaringan.

Prinsip kerja Crude Protein Content:

Metode Crude Protein Content melibatkan pemanasan sampel dalam suasana oksigen, yang menghasilkan gas nitrogen. Gas nitrogen ini kemudian diukur menggunakan detektor, seperti detektor konduktivitas termal, untuk menentukan jumlah nitrogen dalam sampel.

Description

  • Considerable: a 1-disc 120-position autosampler (D200 only) is adopted to directly analyze a large number of samples
  • Quick: the fast cooling function helps the device reach the shutdown temperature in 30 minutes
  • Premium: core pneumatic components of SMC® and mass flow controllers of Bronkhorst® retain the reliability for a long time
  • Cost-effective: new consumables and efficient metal condenser dewatering technologies extend the service life

 

 

Two-stage combustion and oxidation: pure oxygen combustion in the primary combustion tube facilitates preliminary oxidation and digestion of a sample; the remaining part slides through the secondary combustion tube, together with the carrier gas, and is completely digested after being oxidized by the mixture of Pt catalyst and copper oxide;

Three-stage dewatering: most of water is removed by an electronic condenser so that the desiccant can work; traces of water is removed by primary and secondary drying tubes;

Efficient reduction: an efficient reduction agent reduces nitrogen oxides to nitrogen gas and absorbs excess oxygen;

TCD detection: the thermal conductivity detector (TCD) can determine the nitrogen produced by reduction.

Applications

The Dumas Nitrogen/Protein Analyzer is widely used to determine the total nitrogen content in diesel exhaust fluid, cereals (grain), seeds, meat, meat products, animal feed, dairy products, red wine, fertilizers and soil.

 

  1. AOAC uses the Dumas method as a standard method for the determination of crude protein content in grains and oilseeds, meat/meat products and animal feed.
  2. ISO uses the Dumas method for the quality assessment of cereals , pulses, ground cereal products, oilseeds and animal feed.
  3. AACC and ICC use this method to determine grain protein content.
  4. ISO and DIN (German Industrial Standards) also use this method for evaluating dairy products.
  5. AOAC and ISO/DIN use this method as the standard method for the determination of total nitrogen in fertilizers and soils.
  6. ISO, 2015, Natural raw rubber and natural latex Determination of nitrogen content by this method.

Product Features

Sampling System

  1. The 1-disc 120-position autosampler(D200 Only) can conduct a direct analysis of a large sample size
  2. SMC® pneumatic components can inject samples for 100,000 times, with low maintenance costs
  3. The contactless in-situ detection function provided by anti-interference infrared modules can interrupt self-test and pop up messages to avoid maloperation if the sampling disc is occupied (D200 Only)
  4. Users can add samples freely during preparation (either solid or liquid) to enhance experimental efficiency, without stopping the automatic analysis.

Combustion Furnace System

  • High-temperature, pure oxygen, and the two-stage combustion process ensure thorough oxidation and digestion
  • An efficient metal condenser and two-stage desiccants can completely remove water. The desiccants can be reused for 2,000 times.
  • New consumables can reduce nitrogen oxide, absorb excess oxygen, and prolong the service life. Both combustion and reduction tubes can be reused for 1,000 times.

Pneumatic System

  1. A two-stage regulator eliminates the impact of external pressure fluctuations and combustion-induced pressure changes on the detector system to ensure stability
  2. PTFE fittings and SMC® quick couplers allow for clear observation of Contaminants and tool less maintenance
  3. The combustion tube and joint of an original pneumatic connector (patent number: CN106443029B) support easy assembly and disassembly to eliminate air leakage caused by irreversible wear of ferrule compression fittings

Software Function

Windows Chinese Version streamlines operation and improves efficiency. Click to configure parameters and conditions, and control the computer to gather, process, store and print out data.

  1. Wireless communication of balance data: an industrial wireless communication module automatically collects weight data, with the max distance of 100m. The data display window and voice authentication function can adjust value (D200 Only)
  2. Air leakage inspection: the gas path can be checked automatically or manually for the convenience of users
  3. Custom lab report: report format, company name and logo, etc. can be customized
  4. Standby/wake-up function: it reduces the on/off frequency and enhances the experimental efficiency. Carrier gas, power consumption and cost are lowered in standby mode
  5. Intelligent oxygenation function: the software adds oxygen subject to changes in conditions and samples to save reduction agents.
  6. Solution library: a solution library is built in and there’ s no need to develop new approaches
  7. Periodic maintenance reminder: time and cycle of consumable replacement are determined based on sample features, quantity, oxygen consumption, etc., and reminders will be issued timely
  8. Self-diagnosis: if experimental conditions are not up to standard or not prepared, users can choose from 23 fault diagnosis methods to automatically raise a pop-up alarm
  9. Data processing: experimental results can be easily reached by RSD and mean value calculation
  10. Calibration method: accurate experimental data are output by one point or multiple points
  11. Audit trail: the software has developed user permission hierarchy rules for easy traceability in accordance with FDA 21 CFR Part 11(D200 Only)
  12. LIMS: the software share data with the laboratory information management system

Specification:

Working Environment

Power supply

220V AC ±10% 50Hz

Computer

A laptop or a desktop operated with Windows 7 or above and equipped with USB or RS 232 interface

Temperature

5℃~30℃

Humidity

≤85%

 

Technical Parameters

Analysis efficiency

3-4 min each

Weighing accuracy

≤0.1mg

Detection range

0.1-500mg nitrogen

Detection recovery rate

≥99.5%

Relative standard deviation (RSD)

≤0.5% (150mg standard 10% nitrogen)

LOD

0.01mg nitrogen

Sample weight

Solid ≤1g, liquid ≤1mL

Operating system

Windows 7 and above

Sampler

D200:1 disc x 120 1 disc x 40 positions D50:1 disc x 60 positions

Detector

TCD

TCD accuracy

≤0.01℃

Carrier gas

CO2

Furnace temperature

1200℃ (MAX)

Carrier gas purity

99.999%

Oxygen purity

99.999%

Internal module communication mode

RS-485 bus

Peripheral interface

RS232, USB

Correction method

Linear or nonlinear calibration, custom calibration curve

Rated power

2000W

Dimensions (L×W×H)

735mm×560mm×560mm

Net weight

80kg