Aerosol photometer is used to test whether there is leakage on HEPA filter. According to the principle of light scattering, it is portable, yet rugged for in-situ filtration system integrity testing.
The instrument conforms to NSF49 / IEST / ISO14644-3, could realize rapid detection of upstream and downstream concentration detection and real-time display leakage on the host and handheld device, and could find leaking position rapidly and accurately.
APPLICATIONS
Medical facilities and cleanrooms
Biosafety cabinets and fume hoods
Independent filter certifiers
Pharmaceutical manufacturers
STANDARD
> NSF/ANSI 49-2019 Biosafety cabinetry
> ISO14644-3:2005 Cleanrooms and associated controlled environments—Part 3:Test methods
> GB 50073-2013 Code for design of clean workshop
> GB 50591-2010 Code for construction and acceptance of cleanroom
> GMP Factory and device
> YY0569-2005 Biosafety cabinet
> JJF 1815-2020 Calibration specification for class II biosafety cabinet
01. Powerful function > Stable digital photometer. > Dynamic Range: 0.0001μg/L~700μg/L. 02. Good human-computer interaction > 5.0 -inch color screen , touch operation. > Lightweight, portable and equipped with suitcase, easy to carry. > The handheld device is equipped for hard to reach areas, easy to control, display and sample. > Built-in battery((optional)≥3.5H. 03. Data query
> USB & Printer available for real-time reporting. > Audit Tracing & assign permissions to users, further guarantee data integrity. > Sampling data can be imported to PC. 04. Automatic reminder When exceed the set value, light & voice alarm.
> Inform you to replace the high efficiency filter. > Self protection in case of failure.
SPECIFICATION:
Parameter |
Range |
Accuracy |
Sampling flowrate |
28.3L/min |
±2.5% |
Dynamic Range |
(0.0001~700)μg/L |
|
Leakage detection |
0.0001%~100% |
|
Sensitivity |
1% of readings > 0.01% to 100% |
|
Repeatability |
0.5% of readings > 0.01% to 100% |
|
Data storage |
100000 groups |
|
Environment temperature |
(10~35)℃ |
|
Environment humidity |
5%-85%(No condensation、No icing) |
|
Storage requirements |
(-10-40)℃ No condensation when relative humidity lower than 85%. |
|
Aerosol support |
PAO , DOP etc. |
|
Power supply |
AC220V±10% , 50/60Hz |
|
Size |
(length 300×width 330×height 184)mm |
|
Weight |
About 8kg(Battery included) About 14kg(Packing box included) |
|
Power consumption |
<200W |
AEROSOL GENERATOR
When detecting the leakage of high-efficiency filter, you need to cooperate with aerosol generator. It emits aerosol particles with different sizes, and adjusts the aerosol concentration as required to make the upstream concentration reach 10 ~ 20ug / ml. Then the aerosol photometer will detect and display the concentration of the particle mass.
Working pressure:(0-600) kPa
Particle output range:(1.4-56.6) m3/min
Particle concentration :100μg/L (air flow is 5.6 m3/min)
Particle concentration :10μg/L(air flow is 56.6 m3/min)
Dimension:(Length 200×Width 500×Height 280) mm
Aerosol generator is a special instrument for generating DOP aerosol by Laskin nozzle. The embedded adjusting valve could control the working of 4 or 10 nozzles. When the air flow is 1.4m3/min-56.6m3/min, the output aerosol concentration is 10μg/L-100μg/L. The performance of aerosol complies to national standards, The instrument is equipped with aerosol photometer for high efficiency filter leakage detectionand the device is suitable to institutes of medical device testing, centers of disease control, hospitals, pharmaceutical companies, HEPA filter manufacturers for leakage testing of clean rooms and HEPA filters.
The Main Parameters |
Parameter Range |
Resolution |
Maximum Permissible Error (MPE) |
Working pressure |
(0-600) kPa |
1 kPa |
±0.5% |
Particle output range |
(1.4-56.6) m3/min |
||
Particle concentration |
100μg/L (air flow is 5.6 m3/min) |
||
Particle concentration |
10μg/L(air flow is 56.6 m3/min) |
||
Generating method |
4-10 Laskin nozzle |
||
Compressed air |
Embedded compressor |
||
Air type |
Multiple size of particles (Cold generating) |
||
Dimension |
(Length 200×Width 500×Height 280) mm |
||
Noise |
<65dB(A) |
||
Weight |
About 18Kg |
||
Working power |
AC220V±10%,50Hz |
||
Power consumption |
≤500W |