AEROSOL PHOTOMETER, AEROSOL GENERATOR

For leakage detection for HEPA Filter, it is well known to use aerosol photometer for testing. TR-AP6012 Aerosol photometer is a professional testing equipment which is used to test whether there is leakage on HEPA filter.

The downstream of aerosol concentration can be extended to 0.0001 μg/L, and upstream can be extended to 700μg/L.

  • Sampling flowrate:28.3L/min
  • Dynamic Range:(0.0001~700)μg/L
  • Leakage detection:0.0001%~100%
  • Data storage:100000 groups
  • Size:(length 300×width 330×height 184)mm
  • Weight:About 8kg(Battery included) /About 14kg(Packing box included)
  • Power consumption:<200W

Description

Aerosol photometer is used to test whether there is leakage on HEPA filter. According to the principle of light scattering, it is portable, yet rugged for in-situ filtration system integrity testing.
The instrument conforms to NSF49 / IEST / ISO14644-3, could realize rapid detection of upstream and downstream concentration detection and real-time display leakage on the host and handheld device, and could find leaking position rapidly and accurately.
APPLICATIONS
Aerosol photometer (2)

Medical facilities and cleanrooms

Aerosol photometer (3)

Biosafety cabinets and fume hoods

Aerosol photometer (1)

Independent filter certifiers

Aerosol photometer (4)

Pharmaceutical manufacturers

STANDARD

> NSF/ANSI 49-2019     Biosafety cabinetry
>  ISO14644-3:2005       Cleanrooms and associated controlled environments—Part 3:Test methods
>  GB 50073-2013           Code for design of clean workshop
>  GB 50591-2010            Code for construction and acceptance of cleanroom
>  GMP                                Factory and device
>  YY0569-2005              Biosafety cabinet
>  JJF 1815-2020            Calibration specification for class II biosafety cabinet

01. Powerful function
>  Stable digital photometer.
>  Dynamic Range: 0.0001μg/L~700μg/L.

02. Good human-computer interaction
>  5.0 -inch color screen , touch operation.
>  Lightweight, portable and equipped with suitcase, easy to carry.
>  The handheld device is equipped for hard to reach areas, easy to control, display and sample.
>  Built-in battery((optional)≥3.5H.

03. Data query
>  USB & Printer available for real-time reporting.
>  Audit Tracing & assign permissions to users, further guarantee data integrity.
>  Sampling data can be imported to PC.

04. Automatic reminder
When exceed the set value, light & voice alarm.
>  Inform you to replace the high efficiency filter.
>  Self protection in case of failure.

SPECIFICATION:

Parameter

Range

Accuracy

Sampling flowrate

28.3L/min

±2.5%

Dynamic Range

(0.0001~700)μg/L

Leakage detection

0.0001%~100%

Sensitivity

1% of readings > 0.01% to 100%

Repeatability

0.5% of readings > 0.01% to 100%

Data storage

100000 groups

Environment temperature

(10~35)℃

Environment humidity

5%-85%(No condensation、No icing)

Storage requirements

(-10-40)℃

No condensation when relative humidity lower than 85%.

Aerosol support

PAO , DOP etc.

Power supply

AC220V±10% , 50/60Hz

Size

(length 300×width 330×height 184)mm

Weight

About 8kg(Battery included)

About 14kg(Packing box included)

Power consumption

<200W

AEROSOL GENERATOR 
When detecting the leakage of high-efficiency filter, you need to cooperate with aerosol generator. It emits aerosol particles with different sizes, and adjusts the aerosol concentration as required to make the upstream concentration reach 10 ~ 20ug / ml. Then the aerosol photometer will detect and display the concentration of the particle mass.
Working pressure:(0-600) kPa
Particle output range:(1.4-56.6) m3/min
Particle concentration :100μg/L (air flow is 5.6 m3/min)
Particle concentration :10μg/L(air flow is 56.6 m3/min)
Dimension:(Length 200×Width 500×Height 280) mm
Aerosol generator is a special instrument for generating DOP aerosol by Laskin nozzle. The embedded adjusting valve could control the working of 4 or 10 nozzles. When the air flow is 1.4m3/min-56.6m3/min, the output aerosol concentration is 10μg/L-100μg/L. The performance of aerosol complies to national standards, The instrument is equipped with  aerosol photometer for high efficiency filter leakage detectionand the device is suitable to institutes of medical device testing, centers of disease control, hospitals, pharmaceutical companies, HEPA filter manufacturers for leakage testing of clean rooms and HEPA filters.
The Main Parameters
Parameter Range
Resolution
Maximum Permissible Error (MPE)
Working pressure
(0-600) kPa
1 kPa
±0.5%
Particle output range
(1.4-56.6) m3/min
Particle concentration
100μg/L (air flow is 5.6 m3/min)
Particle concentration
10μg/L(air flow is 56.6 m3/min)
Generating method
4-10 Laskin nozzle
Compressed air
Embedded compressor
Air type
Multiple size of particles (Cold generating)
Dimension
(Length 200×Width 500×Height 280) mm
Noise
<65dB(A)
Weight
About 18Kg
Working power
AC220V±10%,50Hz
Power consumption
≤500W