Description
ICP 5100 DUAL VIEW OPTICAL EMISSION SPECTROMETER – INDUCTIVELY COUPLED PLASMA
| PART # | DESCRIPTION | |
| 5100ICP-DV | Fully fast automated Simultaneous system with a Charged Coupled Device Detector (CCD). The Purged
Optical System offers a wide wavelength range high resolution spectrometer with an Echelle Grating and Cross Dispersion Device that provides the FULL Spectrum in a compact area. The RF Plasma Torch is operated from a Solid-State Low Voltage Power Supply. The Plasma view is dual view (combined Radial and Axial views). ICP-Cam provides a color High Resolution continuous view of the Plasma. A Mass Control Module for oxygen addition with Two Port, one for Argon and one for Argon/Oxygen is included. |
|
| FULL SPECIFICATION | ||
| Optical System | : | Echellé Grating Spectrometer (~50grooves/mm) and Prism Cross Dispersion Device. Focal length is 380mm with temperature control for the Spectrometer of 38 oC ±0.1 oC |
| Optical View (Plasma) | : | Dual View (Axial and Radial – Fully Computer Controlled Readout) also Radial and Axial as single Views |
| Wavelength range | : | 165nm – 900nm |
| Detector | Charged Coupled Device (CCD) 1024 x 1024 Pixels. (CCD Pixel Size 24 μm x 24μm) cooled to -360C via a Triple Peltier Cooling System. | |
| Resolution | : | 0.006nm at 200nm |
| RF Plasma System | ||
| Frequency | : | 27.12 MHz. Solid State System. |
| RF Power Range | : | 500 – 1600 Watts (Automatic Power Control with low power standby feature). |
| Tuning | : | Automatic Matching Network |
| Dimensions | : | Small foot print 106(W) x 67(L) x 75(H) cms Weight 180Kg |
| Sample Introduction | : | Cyclonic Spray Chamber
Low Flow Plasma Demountable Torch supplied as standard) Nebulizers – Quartz Concentric, (V-Groove, HF resistant etc. available on request.) |
| Data processing: | : | External Branded PC and instrumental interface Preloaded with ICPWin Software offering FULL control of system and data. Analytical methods available Full program storage with a Library of approx 60,000 Wavelengths. |
| ICP5100DV | : | FULLY AUTOMATED FAST SIMULTANEOUS INDUCTIVELY COUPLED ARGON PLASMA
(ICP) SPECTROMETER with a wavelength range of 160 – 900nm. The Spectrometer has an Echelle grating allowing the simultaneous display of all spectral lines in a single exposure and the analysis of the complete spectrum in a compact area. The Spectrometer is argon purged to allow the analysis of elements in the far UV. The detection system is a Charged Coupled Device (CCD) 1024 x1024 Pixels. (CCD Pixel Size 24 μm x 24μm) with Anti-Blooming protection. High speed acquisition of 500KHz provides a simultaneous Full-Spectrum reading and real-time single pixel sub-array monitoring allowing ultrafast analysis. The “on–board” Solid State RF Generator operates at a frequency of 27.12MHz and has a computer controlled forward power range of 500 – 1600 Watts with real time automatic tuning with stability better than 0.1% and a ECO standby mode. The plasma ignition and generator output is fully monitored via the ICP-Win Software.
The sample introduction system is via a multi- channel (4 Channels) 12 roller peristaltic pump (controlled via the software). A demountable torch, cyclonic spray chamber and a concentric glass nebuliser are supplied as standard (Further options available on request). The torch view is dual i.e. Radial and Axial. ICP5100DV is fully computer controlled via user friendly ICP-Win Software. The multi– window and multi-method analysis program enables fast simultaneous measurements. The software has a built-in library of over 60,000 spectral lines showing inter-element corrections (IEC) and interferences. Visual Background correction points are displayed. The software allows the display of complete “Spectrograms” showing automatically and manually calibrated spectral lines. |
| 5100-001 | ICP5100 24-month operating spares – Comprising of (1x Glass Concentric Nebuliser, 1 x
Quartz Cyclonic Spray Chamber, 1 x Quartz inner & outer Tube, Injector Pump Windings and fittings, ‘O’ Rings) |
|
| 5100-004 | Branded PC Pre-Loaded with ICP-Win Software, with administrative rights (Good
Laboratory Practice) and current Microsoft Windows© operating (Windows 11). System is supplied complete with ICP5100 Interface and Printer. Full Specifications 1. 23.6″ Non-Audio Widescreen Monitor 2. Intel i7-10100 Processor with Graphics 3. DDR4 4Gb 2400 Desktop RAM 4. HP Colou Laser All-In-One Printer 5. Logitech MK120 USB Keyboard & Mouse |
|
| 5100-033 | Multi-Element Tuning and IQ/OQ Qualification Standard 14 Elements in 5% HCl (100ml):
Elements & Concentration (μg/ml) Ba 0.2 Sr 0.2 Mn 5 Al 20 Cd 20 Co 20 Cr 20 Cu 20 Ni 20 Zn 20 Mo 20 K 100 As 100 Se 100 |
|
| 5100-003 | Water Chiller and circulator for cooling to the torch, RF Coil and RF Generator:
Compact 1400-Watt Cooling Capacity. System benefits from an intelligent temperature controller offering ±0.3oC temperature control with alarms for both temperature and flow. 220v AC 50/60Hz · Included as standard with ALL NEW Instruments at NO Charge |
|
| ICP-CAM | ICP-Cam is a colour High Resolution micro camera built into the torch compartment to give a continuous SAFE view of the Plasma on the PC Screen. This allows the operator to monitor the plasma during analysis
· Included as standard with ALL NEW Instruments at NO Charge |
|
| ICP-GCM | Mass Gas Flow control module for oxygen addition with Two Ports, one for Argon and one for Argon/Oxygen is included. The use of oxygen for organic samples eliminates the build up of carbon in the torch and decreases the RSD for the analysis of Oils, Bio-Diesels, Hydro-Carbons, Organics etc.
· Included as standard with ALL NEW Instruments at NO Charge |
|
| 5100-028 | In-Line Sample Filter.
• Reduces the chance of a blocked nebulizer • Protects high throughput valves from damage from particulate matter. • Enhances the day-to-day reproducibility of your results • Purpose-built clog-resistant design • Incorporates a 120 micron filter • Available in 1/16 inch (1.6mm) and 1.3mm OD sample tubing • Filter and ousing Material: HF Resistant PEEK Particle build-up is easily removed by back- flushing and ultrasonic cleaning |
|
| 5100-010 | Argon Nebulizer Gas Humidifier for High TDS
Sample Analysis Argon Nebulizer gas humidifier is designed for use in ICP analyses dealing with high concentrations of dissolved solids. This humidifier utilizes Nafion© tubing that selectively permeates water vapor through its membrane, which humidifies the argon line. The water prevents salt from accumulating in the sample introduction system (nebulizer), which allows for uninterrupted analysis and maintenance free operation. ICP High TDS Sample Analysis · Seawater samples – Oil and gas exploration – Environmental monitoring · Brine samples – Chlor-alkali process Humidifier System · Compact Argon nebulizer gas humidifier · For high TDS samples with concentric glass nebulizer · Improved long-term stability · Improved Levels of Detection. |
| 5100-080 | : | HydraMist Simultaneous Analysis of Hydride Forming Elements and Non Hydride Forming element – HydraMist is a sensitive, cost effective, simple-to-use spray chamber for use with the ICP5100DV that allows simultaneous operation of both conventional pneumatic nebulization and cold-vapour/hydride generation. Cold vapour
generation can provide more than 10 fold improvement in sensitivity on ICP for cold vapour forming elements such As, Sb, Se,Tl and Hg. The HydraMist allows the analysis of BOTH Hydride Forming Elements and Non Hydride Forming Elements simultaneously. System can be interfaced to Auto-Sampler 5100-008 at NO additional costs. |
| 5100-060 | Continuous Flow Dedicated Hydride System with integrated micro-persitaltic pump, 4-Channel System. Specifically designed for ICP5000DV ICP-OES system. The hydride System benefits from a Low-Volume Quartz Gas Liquid Separator. The hydride gasses are separated from liquid waste then transported to the ICP5000 sample Introduction System and directly introduced to the plasma. The system is controlled via ICPHYD-WIN (Pump Network Software) which is included.
The system is designed for the Ultra-Trace Analysis of Mercury (Hg), Arsenic (As), Selenium (Se), Antimony (Sb), Bismuth (Bi), Tellurium (Te) and Tin (Sn). The Hydride System can be fully automated via the PG Instruments ICP5000 Universal Auto-Sampler (5000- 008) at NO additional cost. System is supplied with consumables for 6 months operation. These include Pump Windings, Tubing Adaptors etc. |
|
| 5100-025 | : | HF Sample Introduction system for the ICP5100DV. Comprising of a Demountable Torch
with a 2.0 or 1.8mm Ceramic Injector Tube and a Ceramic Intermediate Tube. A PTFE Cyclonic (double-pass) spray chamber with Helix CT nebulizer interface. The inert PTFE spray chamber has 2 unique features. It has a proprietary Stedi-Flow treatment that ensures the analytical performance of the PTFE spray chamber. · The nebuliser, 5100-023, is made from an Inert material, PEEK and can tolerate a wide variety of solvents including HF · High sensitivity and excellent precision due to concentric design and rigid construction • High tolerance to dissolved solids (typically up to 30%) due to smooth surface of capillary and nebulizer tip · High physical reproducibility: minimal nebulizer to nebulizer variability so that changing nebulizers does not change your analytical method. |
| 5100-023 | : | Nebulizer Routine High Precision for HF
Samples and High Salt Samples. A concentric nebulizer is the most economical nebulizer for high precision analyses involving samples digested in HF. The nebulizer is recommended for samples containing up to 5% HF.: 1. Material: HF Resistant PEEK 2. High physical reproducibility ~ 2% 3. Tolerance to particulates, up to 75μm depending on uptake 4. TDS tolerance, typically up to 30% 5. Low RSD’s due to concentric geometry 6. Uses only standard ICP argon supply pressures 7. Standard 6mm size fits standard cyclonic spray chambers. Adaptors are 8. Standard available uptakes: 0.4 and 1.0 mL/min |
| 5100-021 | :: | Nebuliser for the analysis of HIGH
Concentrations of dissolved solids such as sea water, brines and plating baths etc. Material: Borosilicate glass. High physical reproducibility ~ 1% TDS tolerance, typically up to Tolerance to particulates, typically up to 75μm Low RSD’s due to highly accurate construction Standard available uptake: 2.0 and 0.4mL/min (1mL/min uptake available on request) |
| 5100-008 | : | Random Access X-Y-Z Auto-sampler that is
designed to be sturdy and reliable. The system provide the automated analysis of up to 360 samples and 10 standards, with analysis volumes of 8ml (360 samples), 15ml (240 samples), 20ml (160 samples), 30ml (96 samples MOST COMMON CONFIGURATION) and 50ml (84 samples). A Rinse station is also provided. Interface and software is included for ICP5100DV. Auto-Sampler is supplied with 20 Standard Vials, 240 Sample Vials and Sample Probe |
| EX-001IK | : | Stainless Steel Fume Extraction System
Complete Kit made from High Quality Stainless Steel comprising of: Fan 15mm Diameter (240v AC) with mounting brackets and with Polymer coated corrosion protection blades, Stainless Ducting, Stainless Steel Hood 300mm Diameter, Bends and Adjustable Support Brackets |
| 5100-026 | : | Standard Demountable Torch COMPLETE
designed for ICP5100DV – “PLUG AND PLAY” comprising: Base and Ceramic Inner Tube. Quartz Outer Tube Tapered Quartz Injector 1.5mm Retaining Ring Injector Adapto Injector Ferrule |
| 5100-0261 | : | Quartz Outer Tube for 5100-026 Demountable Torch |
| 5100-0262 | : | Tapered Quartz Injector 1.5mm, OR 2.0mm. for 5100-026 Demountable Torch |
| 5100-0063 | : | Viton O-rings (PKT 10) – for 5100-026 Demountable Torch |
| 5100-0064 | : | Alumina Injector 2.0mm – for 5100-026 Demountable Torch |
| 5100-0065 | : | Alumina Injector 1.0mm – for 5100-026 Demountable Torch |
| 5100-200 | : | PVC Pump Tube 2tag 1.65mm ID Blue/Blue
(PKT 12) for waters and low acidic samples |
| 5100-201 | : | PVC Pump Tube 2tag 1.02mm ID White/White (PKT 12) for waters and low acidic samples. |
| 5100-202 | : | Solva Flex Pump Tube 2tag 1.65mm ID Blue/Blue (PKT 12) for oils and organics |
| 5100-203 | : | Solva Flex Pump Tube 2tag 1.02mm ID White/White (PKT 12) for oils and organics |
| 5100-032 | : | Standard Cyclonic Spray Chamber – Glass Spray Chamber designed for use with the
ICP5100DV standard DEMOUNTABLE ICP torch.(5100-026) |
| 5100-111 | : | RF Work Coil for ICP5100DV 27.12Mhz system |
| 5100-0081 | : | Sample Vials – 16mm x 100mm 14 mL Sample Vials (QTY.1000) |
| 5100-0082 | : | Sample Vials – 30mm x 115mm 50 mL Sample Vials (qty.500) |
| 5100-0083 | : | Auto-Sampler Replacement Probe – Carbon Probe .5mm ID |





WhatsApp us