Description
How Calorimeter Works?
The sample to be analyzed is placed in the decomposition vessel resistant to high pressure. The vessel is placed in the analysis chamber filled with water, which is kept at a constant temperature without being affected by the external environment. The sample in the vessel is ignited by high pressure oxygen (30 bar) and electrodes. With the burning of the combustible sample, the temperature of the vessel increases and it starts to heat the water in the analysis chamber. By measuring the change in water temperature, the calorific value of the sample is calculated.
SPECIFICATION
Temperature Measurement Resolution | 0.0001 K |
Measuring Mode | Isoperibol |
Water Filling and Draining | Automatic |
Ignition | Automatic |
Analysis Duration | < 16 min. |
Preparation Time for Operator | maximum 2 min. |
Range of Measurement | 40 000 J |
Reproducibility | RSD <0,1% |
Operating Oxygen Pressure | 30-40 bar |
Input Power Max. | 200 W |
Rated Voltage | 220V / 120V |
Dimensions | 498x678x396 mm |
Protection Class | III |
Ambient Temperature | 15-35 °C |
Ambient Humidity | < 85% |
Frequency | 50/60 Hz. |
Protection Class According to DIN | IP 20 |
Cooling Medium | Water Circulator |
Water Filter | Included |
Oxygen Filling | Manual |
Degasification | Manual |
Halogen (Decomposition) Vessel | Optional |
Analysis According to | ISO 1928, ISO 18125, ISO 9831, ISO 1716, ASTM D5865, ASTM D5468
ASTM D4809, ASTM D240, ASTM E711, DIN 51 900 |
Interface | USB |
Languages
Water Circulator |
EN/TR
Information |
Operating Options | Heating/Cooling |
Temperature Range | -20 / 100 °C |
Bath Capacity | 7 L |
Temperature Stability | ±0,03 |
Heating Capacity | 1,1 kW |
Cooling Capacity | 0,25 kW |
Power Requirements | 220-240 V / 50-60 Hz |
Dimensions
Software |
292x423x620 mm
Information |
Choosing Analysis Type | Calibration/Standard |
Operating on Sofware While Analysis Continues | Yes |
During Analysis Graphic Progress | Yes |
Analysis Graphic Output | ±0,03 |
Calculation | Gross/ Net Calorific Value |
Excel Output | Yes |
Analysis Control | Device Control/ Analysis Simulation |
File Encryption | Yes |
User Registration | Unlimited |
After Analysis | List and Review Analysis |
Listing | According to: Username Analysis Name, Analysis Date |
Recalculation on Previous Analysis | Yes |
Service Authorization | Full Authorization Log File Access |
User Friendly Service Feedbacks | Warnings and
Auto Problem Identification |
Operating System | Windows |
Reviews