Ion Chromatograph TR-IC160

  1. Pump: Peek
  2. Circuits: BIPOLAR
  3. Detection mode: Single channel
  4. Cations: Already have

Description

FEATURES:
Ion Chromatographic Pump
Pressure range: 0~35 MPa
Volume of flow range: 0.001~9.999 mL/min
Pressure Display accuracy: ≦0.1 Mpa
Volume of flow accuracy (setting error): RSD≦0.1%
Over pressure and leak protection: It will give an alarm and stop working in these situation.

Digital Temperature-control detection system

  1. Conductivity Detector
    Structure: Bipolar Conductivity Detector
    Detection Mode: Double conductivity detection
    Conductance cell size: ≦0.8μl
    Output voltage: -5000~5000 mV (to regulate)
    Measure range: 0~100000 μS/cm (10 level to choose)
    Resolution : ≦0.0020 ns
    Conductance cell temperature: 5~60℃(to regulate)
    Constant temperature accuracy: ±0.01℃
    Baseline noise: ≦0.1%FS
    Baseline drift: ≦0.3%FS
  2. Flow System
    The whole plastic pipeline: made of PEEK materials
    Six-way valve: Rheodyne electromagnetic six-way valve made in America, compression resistance of 7000 psi. Collect signals automatically.
  3. Built-in low-pressure degassing device
    Degree of vacuum: -70kPa
    Inner volume: 300ul
    Organic tolerance: 100%
    Maximum flow rate: 10 mL / min
    Degassing efficiency: degassed at 1.0mL 90%
    Degassing volume: 7.5 mL
  4. Built-in eluent generator
    Type of eluent: KOH / NaOH
    Eluent concentration range: 0.1-100mM / L
    Concentration increment: 0.1 mM
    Flow rate range: 0.1-3.0ml / min
    Maximum operating pressure: 20MPa
    Minimum operating pressure: 5MPa
  5. Analytic Capability
    For high capacity anion column, pH 0~14, compatible organic solvent, etc. can be analyzed at the same time with one injection.
    For high capacity cation column, lycine, ,Choline chloride can be separated and analyzed.
    Different types of detectors can detect several transition metallic ions, heavy metallic ions and special ions.
    qualitative repeatability: ≦0.5% (Cl-, Na+ )
    quantitative repeatability: ≦1.0% (Cl-, Na+)
    Linear range: ≧103
    Minimum concentration of detection: Cl-≤0.0005 ug/ml,BrO3-≤0.001 ug/ml,Li+≤0.001 ug/ml

Reviews

There are no reviews yet.

Be the first to review “Ion Chromatograph TR-IC160”

Your email address will not be published. Required fields are marked *

This site uses Akismet to reduce spam. Learn how your comment data is processed.

WhatsApp WhatsApp us